加工定制否 | 品牌美国CIRS |
型号CIRS 062MQA |
加工定制否 | 品牌美国CIRS |
型号CIRS 062MQA |
美国CIRS 062MQA超声波测试模体,CIRS 062MQA模体简介
The design intent of 062MQA is to provide a tool that can be used for both electron density calibration and image quality assessment of Cone Beam CT systems integrated in radiation therapy devices. The electron calibration function of the phantom enhances the outcome of the adaptive radiation therapy while the image quality features address the fine balance between optimizing image quality while minimizing radiation dose.
The 062MQA CBCT Electron Density & Image Quality Phantom consists of several sections: a 100mm thick body slab, a 100 mm thick CBCT Image Quality phantom, a 50mm thick Electron Density Phantom, and a Bolus consisting of a 37.5mm thick uniform slab, a 12.5mm thick uniform slab and a 50mm thick uniform slab.
The 100 mm thick body section has a central hole that receives the CBCT Image Quality Phantom. Each Bolus slab is drilled to accommodate an ion chamber insert and allow for ion chamber measurements regardless of the position of the Image Quality Insert. The thicknesses of the sections were selected to allow for positioning of any of the layers containing the Image Quality features in the central axis of the beam. Also sections of different thickness decrease the increment with which the electron density section can be offset from the central axis.
The phantom support along with laser alignment marks allow for easy and quick positioning of the phantom on the accelerator couch. The phantom is placed on the support rails with the help of machined grooves. Buffer plates are provided to decrease the artifacts from the support device and to allow the user to access the handles on the support plates. The leveling of the support and implicitly the phantom is done by adjusting leveling feet.
美国CIRS 062MQA特点:
Perform all CT Image QA tests for AAPM TG Report #1
Use ionization chambers for dose measurements
Perform dose measurements using Ion chamber
Calibrate Electron Density in multi-slice CT and Cone Beam CT
Perform central axis and off-set measurements
Position simulated tissue materials in CT & CBCT energy range at 17 different locations
Optimized for volumetric imaging
Quick positioning and customized loading configurations
The One Tool Solution for Electron Density Calibration & Image Quality Assessment
CBCT Image Quality Phantom
The Phantom is comprised of four layers: spatial resolution, CT number linearity/slice thickness, low contrast and uniformity. The positioning of the different layers of the CBCT Image Quality phantom at the central axis can be done with the phantom in the 100mm body section or with the phantom placed directly on the support device.
Spatial Resolution Layer
The Spatial Resolution Layer is designed to evaluate the spatial resolution of IGRT systems. Line pair patterns from 1 lp/cm to 16 lp/cm are embedded in the background. In order to minimize artifacts, each line pair pattern is made from a material with 350HU greater than the background attenuation. The line pair patterns are 3D patterns 12mm in height along the longitudinal axis of the CBCT Image Quality Phantom. The spatial resolution targets are arranged in a circular pattern.
CT Number Linearity and Slice Thickness Layer
The CT number Linearity and Slice Thickness Layer is designed to determine Contrast-to-noise Ratio, Hounsfield number accuracy and Slice Thickness Sensitivity. Six rods made of Air, Low Density Polyethylene (LDPE), Polystyrene, Acrylic, Delrin and Teflon are used to measure the CnR and Hounsfield number Accuracy. Three angled air channels arranged in an equilateral triangle can be used to assess Slice thickness sensitivity.
Low Contrast Layer
The Low Contrast Layer is intended to assess the system's ability to detect small differences in contrast. It contains three sets of low contrast rods with linear attenuation differences of 0.5%, 1% and 2% relative to the background material. The diameters of the low contrast rods were chose to provide a 0.5 ratio between two adjacent rods by cross section and volume. Additional features are designed to evaluate the magnification on the orthogonal axes of the transversal image and as input for calculation of the Point Spread Function and subsequent calculation of Modulation Transfer Function.
Uniformity Layer
The Uniformity Layer is designed to measure the system’s ability to produce uniform images across the field of view of an object with highly homogeneous physical properties in all directions.
Electron Density Phantom
The Phantom consists of nested disks made from Plastic Water?. The nested disks allow simulation of both head and abdomen phantom configurations. Eight different tissue equivalent inserts can be positioned at 17 different locations within the scan field. The geometry of the phantom also enables the user to take measurement offset from the central axis.
NOTE: This product or an optional accessory of this product requires a CIRS dosimetry cavity code before an order can be placed. Please refer to the Dosimetry Cavity Codes document to identify the CIRS code for the probe you intend to use with this product.
深圳市展业达鸿科技有限公司
联系人:吕先生 手机:15920060912 QQ:8329063 邮箱:15920060912@139.com
办公地址:深圳市福田区八卦四路22号华晟达大厦7楼***
企业类型 | 有限责任公司(自然人投资或控股) | 统一社会信用代码 | 914403005977598362 |
---|---|---|---|
成立日期 | 2012-06-08 | 法定代表人/负责人 | 吕宏齐 |
注册资本 | 500万(元) | 注册地址 | 深圳市福田区园岭街道鹏盛社区八卦一路八卦岭工业区619栋628 |
营业期限 | 2012-06-08 至 无固定期限 | 登记机关 | 深圳市市场监督管理局福田监管局 |
经营范围 | 仪器仪表、环境环保仪器、一类医疗器械的技术开发、销售及上门维修;电子仪器、医疗设备、医疗测试技术开发、技术咨询;软件的技术咨询、技术服务\技术开发,投资兴办实业(具体项目另行申报),国内贸易;货物及技术进出口。(象牙及其制品除外,法律、行政法规、国务院决定禁止的项目除外,限制的项目须取得许可后方可经营);仪器仪表制造;电子元器件制造;计算机软硬件及外围设备制造。(除依法须经批准的项目外,凭营业执照依法自主开展经营活动)^二类、三类医疗器械的的技术开发、销售及上门维修 |